A semi-automated volumetric software for segmentation and perfusion parameter quantification of brain tumors using 320-row multidetector computed tomography: a validation study

Soo Young Chae, Sangil Suh, Inseon Ryoo, Arim Park, Kyoung Jin Noh, Hackjoon Shim, Hae Young Seol

Research output: Contribution to journalArticlepeer-review

7 Citations (Scopus)

Abstract

Purpose: We developed a semi-automated volumetric software, NPerfusion, to segment brain tumors and quantify perfusion parameters on whole-brain CT perfusion (WBCTP) images. The purpose of this study was to assess the feasibility of the software and to validate its performance compared with manual segmentation. Methods: Twenty-nine patients with pathologically proven brain tumors who underwent preoperative WBCTP between August 2012 and February 2015 were included. Three perfusion parameters, arterial flow (AF), equivalent blood volume (EBV), and Patlak flow (PF, which is a measure of permeability of capillaries), of brain tumors were generated by a commercial software and then quantified volumetrically by NPerfusion, which also semi-automatically segmented tumor boundaries. The quantification was validated by comparison with that of manual segmentation in terms of the concordance correlation coefficient and Bland-Altman analysis. Results: With NPerfusion, we successfully performed segmentation and quantified whole volumetric perfusion parameters of all 29 brain tumors that showed consistent perfusion trends with previous studies. The validation of the perfusion parameter quantification exhibited almost perfect agreement with manual segmentation, with Lin concordance correlation coefficients (ρc) for AF, EBV, and PF of 0.9988, 0.9994, and 0.9976, respectively. On Bland-Altman analysis, most differences between this software and manual segmentation on the commercial software were within the limit of agreement. Conclusions: NPerfusion successfully performs segmentation of brain tumors and calculates perfusion parameters of brain tumors. We validated this semi-automated segmentation software by comparing it with manual segmentation. NPerfusion can be used to calculate volumetric perfusion parameters of brain tumors from WBCTP.

Original languageEnglish
Pages (from-to)461-469
Number of pages9
JournalNeuroradiology
Volume59
Issue number5
DOIs
Publication statusPublished - 2017 May 1
Externally publishedYes

Bibliographical note

Publisher Copyright:
© 2017, Springer-Verlag Berlin Heidelberg.

Keywords

  • Brain tumor
  • CT perfusion
  • Segmentation
  • Semi-automated segmentation
  • Whole-brain CT perfusion

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Clinical Neurology
  • Cardiology and Cardiovascular Medicine