Abstract
A novel high-throughput second-order correlation measurement system is developed that records and makes use of all the arrival times of photons detected at both start and stop detectors. This system is suitable, particularly for a light source having a high photon flux and a long coherence time, since it is more efficient than conventional methods by an amount equal to the product of the count rate and the correlation time of the light source. We have used this system in carefully investigating the dead time effects of detectors and photon counters on the second-order correlation function in the two-detector configuration. For a nonstationary light source, a distortion of the original signal was observed at high photon flux. A systematic way of calibrating the second-order correlation function has been devised by introducing the concept of an effective dead time of the entire measurement system.