T1 - Comparative evaluation of Plateletworks, Multiplate analyzer and Platelet function analyzer-200 in cardiology patients
AU - Kim, Jeeyong
AU - Cho, Chi Hyun
AU - Jung, Bo Kyeung
AU - Nam, Jeonghun
AU - Seo, Hong Seog
AU - Shin, Sehyun
AU - Lim, Chae Seung
N1 - Funding Information:
This study was supported by a grant from the Korea Health Technology R&D Project, Ministry of Health & Welfare, Republic of Korea (Grant No. HI14C0670).
Publisher Copyright:
© 2018 - IOS Press and the authors. All rights reserved.
PY - 2018
Y1 - 2018
N2 - The objective of this study was to comparatively evaluate three commercial whole-blood platelet function analyzer systems: Platelet Function Analyzer-200 (PFA; Siemens Canada, Mississauga, Ontario, Canada), Multiplate analyzer (MP; Roche Diagnostics International Ltd., Rotkreuz, Switzerland), and Plateletworks Combo-25 kit (PLW; Helena Laboratories, Beaumont, TX, USA). Venipuncture was performed on 160 patients who visited a department of cardiology. Pairwise agreement among the three platelet function assays was assessed using Cohen's kappa coefficient and percent agreement within the reference limit. Kappa values with the same agonists were poor between PFA-collagen (COL; agonist)/adenosine diphosphate (ADP) and MP-ADP (-0.147), PFA-COL/ADP and PLW-ADP (0.089), MP-ADP and PLW-ADP (0.039), PFA-COL/ADP and MP-COL (-0.039), and between PFA-COL/ADP and PLW-COL (-0.067). Nonetheless, kappa values for the same assay principle with a different agonist were slightly higher between PFA-COL/ADP and PFA-COL/EPI (0.352), MP-ADP and MP-COL (0.235), and between PLW-ADP and PLW-COL (0.247). The range of percent agreement values was 38.7% to 73.8%. Therefore, various measurements of platelet function by more than one method were needed to obtain a reliable interpretation of platelet function considering low kappa coefficient and modest percent agreement rates among 3 different platelet function tests.
AB - The objective of this study was to comparatively evaluate three commercial whole-blood platelet function analyzer systems: Platelet Function Analyzer-200 (PFA; Siemens Canada, Mississauga, Ontario, Canada), Multiplate analyzer (MP; Roche Diagnostics International Ltd., Rotkreuz, Switzerland), and Plateletworks Combo-25 kit (PLW; Helena Laboratories, Beaumont, TX, USA). Venipuncture was performed on 160 patients who visited a department of cardiology. Pairwise agreement among the three platelet function assays was assessed using Cohen's kappa coefficient and percent agreement within the reference limit. Kappa values with the same agonists were poor between PFA-collagen (COL; agonist)/adenosine diphosphate (ADP) and MP-ADP (-0.147), PFA-COL/ADP and PLW-ADP (0.089), MP-ADP and PLW-ADP (0.039), PFA-COL/ADP and MP-COL (-0.039), and between PFA-COL/ADP and PLW-COL (-0.067). Nonetheless, kappa values for the same assay principle with a different agonist were slightly higher between PFA-COL/ADP and PFA-COL/EPI (0.352), MP-ADP and MP-COL (0.235), and between PLW-ADP and PLW-COL (0.247). The range of percent agreement values was 38.7% to 73.8%. Therefore, various measurements of platelet function by more than one method were needed to obtain a reliable interpretation of platelet function considering low kappa coefficient and modest percent agreement rates among 3 different platelet function tests.
KW - Multiplate analyzer
KW - Platelet function analyzer-200
KW - Plateletworks
UR - http://www.scopus.com/inward/record.url?scp=85057779869&partnerID=8YFLogxK
U2 - 10.3233/CH-170331
DO - 10.3233/CH-170331
M3 - Article
C2 - 29710682
AN - SCOPUS:85057779869
SN - 1386-0291
VL - 70
SP - 257
EP - 265
JO - Clinical Hemorheology and Microcirculation
JF - Clinical Hemorheology and Microcirculation
IS - 3
ER -