Development of a multi-channel waveform-sampling front-end ASIC and DOI detector for APD based animal PET

J. Y. Yeom, H. Takahashi, T. Ishitsu, M. Nakazawa, H. Murayama

Research output: Contribution to journalConference articlepeer-review

2 Citations (Scopus)

Abstract

A new multi-channel Waveform Sampling Front-End (WSFE) Application Specific Integrated Circuit (ASIC) for Positron Emission Tomography (PET) has been developed to digitize signals from scintillating crystals at an early stage to suppress noise while facilitating signal processing. Each channel of the chip consists of a preamplifier, a variable gain amplifier (VGA) and a fast Analog to Digital Converter (ADC). This multi-channel chip will be used to readout each pixel of a phoswich detector based on GSO scintillators coupled individually to a multi-array Avalanche Photodiode (APD). Two WSFE ASIC have been developed together with some low noise preamplifiers in a separate chip.

Original languageEnglish
Article numberM3-16
Pages (from-to)1794-1796
Number of pages3
JournalIEEE Nuclear Science Symposium Conference Record
Volume3
Publication statusPublished - 2003
Externally publishedYes
Event2003 IEEE Nuclear Science Symposium Conference Record - Nuclear Science Symposium, Medical Imaging Conference - Portland, OR, United States
Duration: 2003 Oct 192003 Oct 25

ASJC Scopus subject areas

  • Radiation
  • Nuclear and High Energy Physics
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Development of a multi-channel waveform-sampling front-end ASIC and DOI detector for APD based animal PET'. Together they form a unique fingerprint.