Abstract
Objective: The objective of the present study was to compare and evaluate the accuracy of three-dimensional (3D) image data acquired from cone beam computed tomography (CBCT) and a dental scanner using 3D software. Materials and methods: After selecting the full-arch forms of the maxilla and mandible as the master cast, the master cast was scanned via a high-precision optical scanner for use as master model data. The model was scanned 12 times each using CBCT and a dental scanner. Scanned data were superimposed onto the master cast data for evaluation of accuracy and repeatability. Results: Although significant differences in both accuracy and repeatability were seen between CBCT and dental scanner (P < 0.05), repeatability of the maxillary arch showed little difference, with CBCT and scanner having values of 17 ± 2 μm and 22 ± 5 μm, respectively. Meanwhile, repeatability of the mandibular arch with CBCT and scanner was 15 ± 0 μm and 19 ± 3 μm, respectively. Since good repeatability was shown, this demonstrated that data can be stably acquired. Conclusions: The present study demonstrated the feasibility of using a dental scanner to create a digital model as a substitute for a plaster model for use in orthodontic diagnosis and device fabrication.
Original language | English |
---|---|
Pages (from-to) | 65-73 |
Number of pages | 9 |
Journal | International Journal of Computerized Dentistry |
Volume | 20 |
Issue number | 1 |
Publication status | Published - 2017 |
Keywords
- 3D software
- Accuracy
- Cone beam computed tomography
- Dental scanner
- Digital model
- Repeatability
ASJC Scopus subject areas
- Medicine(all)