Abstract
BACKGROUND: Infants admitted to neonate intensive care units (NICUs) are placed in incubators to maintain body temperature and condition, which undergo normal radiographs and are exposed to radiation. Furthermore, different incubator structures in different hospitals exhibit varying object to image receptor distance (OID), source to image receptor distance (SID), presence of canopy, which results in variations in X-ray radiation conditions and doses absorbed by the neonatal patients. OBJECTIVE: To measure organ dose exposed to neonatal patient in different incubator settings. METHODS: A portable X-ray was performed on a neonatal patient placed in an incubator to identify disease progress, the injection path of the drug, and various factors. To minimize direct contact between neonatal patients and image receptor, radiologic technologists place the image receptor on a tray underneath the incubator and place the portable X-ray tube on top of the acrylic canopy of the incubators. SID and OID settings and value of organ dose exposed to the patient varied based on the incubator structure, and the organ absorbed dose was determined using Monte Carlo N-Particle (MCNP) simulation, PC-based Monte Carlo program (PCXMC) 2.0 simulation, and neonate phantoms. RESULTS: Evaluations of organ dose of neonatal patients in three hospitals with different incubator settings reveal that the average organ dose differs by 36% depending on change in OID and SID settings and reduces by 10% with an acrylic canopy. Therefore, owing to the presence of an acrylic canopy on the top of the incubator and the longer SID with the corresponding shorter OID, a lower dose was absorbed by organs of neonatal patient. CONCLUSION: Our results provide proof that proper incubator standard decreases organ dose to neonatal patient during continuously diagnostic X-ray procedure.
Original language | English |
---|---|
Pages (from-to) | 333-342 |
Number of pages | 10 |
Journal | Journal of X-Ray Science and Technology |
Volume | 30 |
Issue number | 2 |
DOIs | |
Publication status | Published - 2022 |
Bibliographical note
Funding Information:This work was supported by the Korea Medical Device Development Fund grant funded by the Korean government (the Ministry of Science and ICT, the Ministry of Trade, Industry and Energy, the Ministry of Health and Welfare, the Ministry of Food and Drug Safety) (Project Number: KMDF PR 20200901 0262). This work was also supported by the BB21+ Project in 2021.
Publisher Copyright:
© 2022 - IOS Press. All rights reserved.
Keywords
- Monte Carlo simulation
- Neonatal patient
- Object to image receptor distance
- Organ dose estimation
- X-ray source to image receptor distance
ASJC Scopus subject areas
- Radiation
- Instrumentation
- Radiology Nuclear Medicine and imaging
- Condensed Matter Physics
- Electrical and Electronic Engineering