Abstract
This paper describes two general methods: field map processing and NMR lineshape analysis, that we are currently using for the on-going research of field improvement techniques specifically applied to our 700 MHz LTS/HTS NMR magnet. The validity of the methods is verified with comparison between calculations and experiments. Also, a simple but effective analysis has been used to identify the principal source of a remanent magnetic field measured in the bore of the 700 MHz LTS/HTS NMR magnet: the Screening Current induced Field (SCF) from the 100-MHz HTS insert, comprised of 48 double pancake coils, each wound with Bi2223 tape.
Original language | English |
---|---|
Article number | 4497933 |
Pages (from-to) | 856-859 |
Number of pages | 4 |
Journal | IEEE Transactions on Applied Superconductivity |
Volume | 18 |
Issue number | 2 |
DOIs | |
Publication status | Published - 2008 Jun |
Keywords
- Field mapping
- LTS/HTS NMR magnet
- Nmr lineshape
- Screening currents induced field (SCF)
ASJC Scopus subject areas
- Electronic, Optical and Magnetic Materials
- Condensed Matter Physics