Abstract
The purpose of this study is to develop a software for optimized registering of film dosimetry with calculated dose for Intensity Modulated Radiation Therapy (IMRT) quality assurance. The newly developed software consists of a film device driver and the verification of fluence map. A film device driver follows the TWAIN 32 protocol. This driver enables us to easily calibrate and scan the films in any film scanner. The fluence map verification software is programmed by C++, which is suitable for the extension of future work. Taking advantages of that multi platform Qt (Trolltech, Norway) GUI library, we have made it run on Windows and Linux operating systems even Mac OSX. For drawing and viewing the fluence map in 2 and 3D, OpenGL (SGI Corp., USA) is implemented. Both absolute point dosimetry and 3D film dosimetry were performed throughout the IMRT using Clinac 21EX’s millennium MLC. With these custom-written software modules, the measured and calculated dose distributions for axial planes were superimposed by coincidence of their origins, followed by comparison of the point doses at all matched positions. Then, with the optimization algorithm the setup errors were recovered. We have developed customwritten software modules for managing dose-distribution files, optimizing the position, calculating dose differences, and printing out the report. Differences between the calculated and measured doses over the 3% criterion could be reduced by 15% after applying the optimization algorithm. Optimization dramatically reduced the difference between measured and calculated dose distributions in all cases investigated.
Original language | English |
---|---|
Title of host publication | IFMBE Proceedings |
Editors | Sun I. Kim, Tae Suk Suh |
Publisher | Springer Verlag |
Pages | 2000-2003 |
Number of pages | 4 |
Edition | 1 |
ISBN (Print) | 9783540368397 |
DOIs | |
Publication status | Published - 2007 |
Externally published | Yes |
Event | 10th World Congress on Medical Physics and Biomedical Engineering, WC 2006 - Seoul, Korea, Republic of Duration: 2006 Aug 27 → 2006 Sept 1 |
Publication series
Name | IFMBE Proceedings |
---|---|
Number | 1 |
Volume | 14 |
ISSN (Print) | 1680-0737 |
ISSN (Electronic) | 1433-9277 |
Other
Other | 10th World Congress on Medical Physics and Biomedical Engineering, WC 2006 |
---|---|
Country/Territory | Korea, Republic of |
City | Seoul |
Period | 06/8/27 → 06/9/1 |
Bibliographical note
Publisher Copyright:© International Federation for Medical and Biological Engineering 2007.
Keywords
- Film dosimetry
- Intensity modulated radiation therapy
- Optimization algorithm
- Quality assurance
ASJC Scopus subject areas
- Bioengineering
- Biomedical Engineering