Performance Evaluation of Compton Micro-PET for Detector Modalities: A Monte Carlo Study

Changyeon Yoon, Wonho Lee

Research output: Contribution to journalArticlepeer-review

Abstract

Performance of Compton positron emission tomography (PET) is studied in this paper using qualitative and quantitative methods. Lutetium-yttrium oxyorthosilicate (LYSO), lutetium-gadolinium oxyorthosilicate (LGSO), and CdZnTe (CZT) materials are used for Compton PET. LYSO is widely used for conventional PET, and LGSO is a prospective scintillator material for PET detectors. CZT is one of the semiconductor materials that have high energy and position resolution. For conventional PET, only the photoelectric effect is considered a valid interaction for image reconstruction. However, Compton scattering tracing technology is applied for our Compton PET to additionally use Compton scattering events for image reconstruction. It is relatively difficult to use multiple layers for PET made of scintillators, as electronic circuits must be attached to each layer. For this reason, conventional PET generally uses only one layer for each detector module and limits the spatial resolution in the depth direction. In contrast, it is possible for a CZT detector to measure a depth of interest based on the cathode-to-anode signal ratio or electron drift time with relatively simple electronic circuits. Furthermore, CZT materials have high spatial and energy resolutions. Therefore, the position and energy information of the radiation interactions in the detector module can be precisely calculated to determine the interaction sequence, and hence, the information from the Compton scattering can be used for image reconstruction in PET. For this reason, the reconstructed image of CZT PET can show better quality than those of scintillator PETs. The detection efficiency and quality of the reconstructed image are significantly increased by including the Compton scattering effect as a valid interaction process for image reconstruction because Compton scattering has twice the interaction probability of the photoelectric effect at 511 keV. In this paper, the effectiveness of including Compton scattering events for PET reconstruction was evaluated for scintillators and CZT semiconductor detectors. The maximum likelihood expectation and maximization reconstruction method was applied for conventional and Compton PET reconstruction, and the qualities of the reconstructed images were evaluated.

Original languageEnglish
Pages (from-to)386-395
Number of pages10
JournalNuclear Technology
Volume204
Issue number3
DOIs
Publication statusPublished - 2018 Dec 2

Bibliographical note

Funding Information:
This work was supported by the Nuclear Safety Research Program through the Korea Foundation of Nuclear Safety and granted financial resources from the Nuclear Safety and Security Commission, Korea (1603015) and Korea University (K1808611).

Publisher Copyright:
© 2018, © 2018 American Nuclear Society.

Keywords

  • Compton PET
  • Monte Carlo study
  • semiconductor PET

ASJC Scopus subject areas

  • Nuclear and High Energy Physics
  • Nuclear Energy and Engineering
  • Condensed Matter Physics

Fingerprint

Dive into the research topics of 'Performance Evaluation of Compton Micro-PET for Detector Modalities: A Monte Carlo Study'. Together they form a unique fingerprint.

Cite this