Abstract
To quantify the differences between heterogeneous dose estimates from the calculation algorithm of 3-D treatment planning systems and dosimetry measurements, radiation was delivered in accordance with IMRT and dynamic conformal arc radiotherapy (DCART) QC procedures using multiple heterogeneous head & neck phantoms. The multiple inhomogeneous head and neck phantoms were constructed by various materials such as water equivalent homogeneous (acryl), air equivalent (cavity), bone equivalent (teflon) materials. The absolute and relative dosimetry were done using pinpoint ionization chamber, film and TLD (Harshaw 100). With custom-written software modules, the measured and calculated dose distributions were superimposed and compared. The point dose measured at the interface between water and air material region shows about 7% mismatch with the treatment planning system in IMRT. This dose difference is even increased at the interface between bone and water material region reaches up to 18% under-dose comparative to planned value. Our results suggest that the effectiveness of the imhomogeneity corrections used in IMRT and DCART planning should be evaluated in order to ensure meaningful quality assurance and delivery.
Original language | English |
---|---|
Title of host publication | IFMBE Proceedings |
Editors | Sun I. Kim, Tae Suk Suh |
Publisher | Springer Verlag |
Pages | 1867-1870 |
Number of pages | 4 |
Edition | 1 |
ISBN (Print) | 9783540368397 |
DOIs | |
Publication status | Published - 2007 |
Event | 10th World Congress on Medical Physics and Biomedical Engineering, WC 2006 - Seoul, Korea, Republic of Duration: 2006 Aug 27 → 2006 Sept 1 |
Publication series
Name | IFMBE Proceedings |
---|---|
Number | 1 |
Volume | 14 |
ISSN (Print) | 1680-0737 |
ISSN (Electronic) | 1433-9277 |
Other
Other | 10th World Congress on Medical Physics and Biomedical Engineering, WC 2006 |
---|---|
Country/Territory | Korea, Republic of |
City | Seoul |
Period | 06/8/27 → 06/9/1 |
Bibliographical note
Publisher Copyright:© International Federation for Medical and Biological Engineering 2007.
Keywords
- Conformal dynamic arc radiotherapy
- Head and neck phantom
- Inhomogeniety
ASJC Scopus subject areas
- Bioengineering
- Biomedical Engineering