Quality assurance of inhomogeneity correction algorithm in the intensity modulated radiotherapy and dynamic conformal arc radiotherapy using multiple heterogeneous head and neck phantoms

Dongho Shin, Myonggeun Yoon, Sung Yong Park, Suk Lee, Sang Hoon Lee, Hyun Do Huh, Dae Yong Kim, Kwan Ho Cho, Dong Ho Shin

    Research output: Chapter in Book/Report/Conference proceedingConference contribution

    Abstract

    To quantify the differences between heterogeneous dose estimates from the calculation algorithm of 3-D treatment planning systems and dosimetry measurements, radiation was delivered in accordance with IMRT and dynamic conformal arc radiotherapy (DCART) QC procedures using multiple heterogeneous head & neck phantoms. The multiple inhomogeneous head and neck phantoms were constructed by various materials such as water equivalent homogeneous (acryl), air equivalent (cavity), bone equivalent (teflon) materials. The absolute and relative dosimetry were done using pinpoint ionization chamber, film and TLD (Harshaw 100). With custom-written software modules, the measured and calculated dose distributions were superimposed and compared. The point dose measured at the interface between water and air material region shows about 7% mismatch with the treatment planning system in IMRT. This dose difference is even increased at the interface between bone and water material region reaches up to 18% under-dose comparative to planned value. Our results suggest that the effectiveness of the imhomogeneity corrections used in IMRT and DCART planning should be evaluated in order to ensure meaningful quality assurance and delivery.

    Original languageEnglish
    Title of host publicationIFMBE Proceedings
    EditorsSun I. Kim, Tae Suk Suh
    PublisherSpringer Verlag
    Pages1867-1870
    Number of pages4
    Edition1
    ISBN (Print)9783540368397
    DOIs
    Publication statusPublished - 2007
    Event10th World Congress on Medical Physics and Biomedical Engineering, WC 2006 - Seoul, Korea, Republic of
    Duration: 2006 Aug 272006 Sept 1

    Publication series

    NameIFMBE Proceedings
    Number1
    Volume14
    ISSN (Print)1680-0737
    ISSN (Electronic)1433-9277

    Other

    Other10th World Congress on Medical Physics and Biomedical Engineering, WC 2006
    Country/TerritoryKorea, Republic of
    CitySeoul
    Period06/8/2706/9/1

    Bibliographical note

    Publisher Copyright:
    © International Federation for Medical and Biological Engineering 2007.

    Keywords

    • Conformal dynamic arc radiotherapy
    • Head and neck phantom
    • Inhomogeniety

    ASJC Scopus subject areas

    • Bioengineering
    • Biomedical Engineering

    Fingerprint

    Dive into the research topics of 'Quality assurance of inhomogeneity correction algorithm in the intensity modulated radiotherapy and dynamic conformal arc radiotherapy using multiple heterogeneous head and neck phantoms'. Together they form a unique fingerprint.

    Cite this