TY - JOUR
T1 - Simulation study of a novel target oriented SPECT design using a variable pinhole collimator
AU - Bae, Seungbin
AU - Chun, Jaehee
AU - Cha, Hyemi
AU - Yeom, Jung Yeol
AU - Lee, Kisung
AU - Lee, Hakjae
N1 - Funding Information:
This research was supported by the National R&D Program through the Korea Institute of Radiological and Medical Sciences grant (No.1711021927) and National Research Foundation of Korea (NRF) grant (No.2016R1A2B2007551 and 2016R1A6A3A11931787) funded by the Ministry of Science, ICT & Future Planning.
Publisher Copyright:
© 2016 American Association of Physicists in Medicine.
PY - 2017/2
Y1 - 2017/2
N2 - Purpose: In the past decade, demands for organ specific (target oriented) single-photon emission computed tomography (SPECT) is increasing, and several groups have conducted studies on developing clinical dedicated SPECT with pinhole collimator to improve the spatial resolution. However, acceptance angle of the collimator cannot be adjusted to fit the different ROIs of target objects because the shape of pinhole could not be changed, and the magnifying factor cannot be maximized as the collimator-to-detector distance is fixed. Furthermore, those dedicated pinhole SPECTs are typically made for a single purpose and therefore possess a drawback in that it cannot be utilized for any other purpose. In this study, we propose a novel SPECT system using variable pinhole collimator (VP SPECT) whose parameters are flexible. Methods: The proposed variable pinhole collimator is modeled on conventional pinhole by piling several tungsten layers of different apertures. Depending on the combination of the holes in each layer, a variety of hole diameters and acceptance angles of the pinhole can be made. In addition, VP SPECT system allows attaching the collimator to the object as close as possible to maximize the sensitivity and adjust the distance of the pinhole from the scintillation detector to optimize the system resolution for each rotation angle, automatically. For quantitative measurement, we compared the sensitivity and spatial resolution of VP SPECT with those of conventional pinhole SPECT. To determine the possibility of the clinical and preclinical use of proposed system, a digital mouse whole-body (MOBY) phantom is used for simulating the live mouse model. Results: The result of simulation using ultra-micro hot spot phantom shows that the sensitivity of the proposed VP SPECT system is about 297% of that of the conventional system. While hot rods of diameter 0.6 mm can be distinguished in the image with the proposed VP SPECT system, 1.2-mm hot rods are barely discernible in the conventional pinhole SPECT image. According to the result of MOBY phantom simulation, heart walls separated by 3 mm were not distinguished in conventional pinhole SPECT images, but were clearly discerned in VP SPECT images. Conclusions: In this study, we designed a novel pinhole collimator for SPECT and presented preliminary results of target oriented imaging with a simulation study. Currently, we are pursuing strategies to realize the proposed system, with the goal to apply the technology into a high-sensitivity and high-resolution preclinical SPECT. Should VP SPECT be applied to the clinical setting, we anticipate a high-sensitivity, high-resolution system for applications such as heart dedicated SPECT or related fields.
AB - Purpose: In the past decade, demands for organ specific (target oriented) single-photon emission computed tomography (SPECT) is increasing, and several groups have conducted studies on developing clinical dedicated SPECT with pinhole collimator to improve the spatial resolution. However, acceptance angle of the collimator cannot be adjusted to fit the different ROIs of target objects because the shape of pinhole could not be changed, and the magnifying factor cannot be maximized as the collimator-to-detector distance is fixed. Furthermore, those dedicated pinhole SPECTs are typically made for a single purpose and therefore possess a drawback in that it cannot be utilized for any other purpose. In this study, we propose a novel SPECT system using variable pinhole collimator (VP SPECT) whose parameters are flexible. Methods: The proposed variable pinhole collimator is modeled on conventional pinhole by piling several tungsten layers of different apertures. Depending on the combination of the holes in each layer, a variety of hole diameters and acceptance angles of the pinhole can be made. In addition, VP SPECT system allows attaching the collimator to the object as close as possible to maximize the sensitivity and adjust the distance of the pinhole from the scintillation detector to optimize the system resolution for each rotation angle, automatically. For quantitative measurement, we compared the sensitivity and spatial resolution of VP SPECT with those of conventional pinhole SPECT. To determine the possibility of the clinical and preclinical use of proposed system, a digital mouse whole-body (MOBY) phantom is used for simulating the live mouse model. Results: The result of simulation using ultra-micro hot spot phantom shows that the sensitivity of the proposed VP SPECT system is about 297% of that of the conventional system. While hot rods of diameter 0.6 mm can be distinguished in the image with the proposed VP SPECT system, 1.2-mm hot rods are barely discernible in the conventional pinhole SPECT image. According to the result of MOBY phantom simulation, heart walls separated by 3 mm were not distinguished in conventional pinhole SPECT images, but were clearly discerned in VP SPECT images. Conclusions: In this study, we designed a novel pinhole collimator for SPECT and presented preliminary results of target oriented imaging with a simulation study. Currently, we are pursuing strategies to realize the proposed system, with the goal to apply the technology into a high-sensitivity and high-resolution preclinical SPECT. Should VP SPECT be applied to the clinical setting, we anticipate a high-sensitivity, high-resolution system for applications such as heart dedicated SPECT or related fields.
KW - pinhole collimator
KW - simulation
KW - single-photon emission computed tomography
UR - http://www.scopus.com/inward/record.url?scp=85015599626&partnerID=8YFLogxK
U2 - 10.1002/mp.12075
DO - 10.1002/mp.12075
M3 - Article
C2 - 28032904
AN - SCOPUS:85015599626
SN - 0094-2405
VL - 44
SP - 470
EP - 478
JO - Medical physics
JF - Medical physics
IS - 2
ER -