    This study applied statistical process control to set and verify the quality assurances (QA) tolerance standard for our hospital’s characteristics with the criteria standards that are applied to all the treatment sites with this analysis. Gamma test factor of delivery quality assurances (DQA) was based on 3%/3 mm. Head and neck, breast, prostate cases of intensity modulated radiation therapy (IMRT) or volumetric arc radiation therapy (VMAT) were selected for the analysis of the QA treatment sites. The numbers of data used in the analysis were 73 and 68 for head and neck patients. Prostate and breast were 49 and 152 by MapCHECK and ArcCHECK respectively. Cp value of head and neck and prostate QA were above 1.0, Cpml is 1.53 and 1.71 respectively, which is close to the target value of 100%. Cpml value of breast (IMRT) was 1.67, data values are close to the target value of 95%. But value of was 0.90, which means that the data values are widely distributed. Cp and Cpml of breast VMAT QA were respectively 1.07 and 2.10. This suggests that the VMAT QA has better process capability than the IMRT QA. Consequently, we should pay more attention to planning and QA before treatment for breast Radiotherapy.

    Original languageEnglish
    Pages (from-to)717-721
    Number of pages5
    JournalJournal of the Korean Physical Society
    Volume71
    Issue number10
    DOIs
    Publication statusPublished - 2017 Nov 1

    Bibliographical note

    Funding Information:
    This research was supported by Basic Science Research Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Education (No. 2015R1D1A1A01060463) and the Nuclear Safety Research Program through the Korea Radiation Safety Foundation(KORSAFe) and the Nuclear Safety and Security Commission(NSSC), Republic of Korea (Grant No. 1305033).

    Publisher Copyright:
    © 2017, The Korean Physical Society.

    Keywords

    • Gamma index
    • Gamma pass rate
    • IMRT/VMAT QA
    • Process capability index
    • Statistical process control

    ASJC Scopus subject areas

    • General Physics and Astronomy

    Fingerprint

    Dive into the research topics of 'Statistical process control analysis for patient quality assurance of intensity modulated radiation therapy'. Together they form a unique fingerprint.

    Cite this